Fully Automatic Biochemistry Analyzer - Mispa CXL Pro Plus
Mispa CXL Pro Plus is a fully automated, discrete, random access clinical Chemistry analyzer. Mispa CXL Pro Plus offers 240 tests per hour. Mispa CXL Pro Plus has a Holographic Concave Flat Field Grating (HCFG) rear spectrophotometry system for best reliability and maximum accuracy. CXL Pro Plus comes with Special Probes, having an internal water pressure wash and external waterfall wash, eliminating carryover. Mispa CXL Pro plus is a modern analyzer, able to perform endpoint, fixed time, kinetic, single & dual reagents chemistry mono & bi chromatic, linear & non-linear multipoint calibration. Mispa CXL Pro Plus is an all-in-one solution for clinical chemistry analysis.
Features
Triple speed mixing - (ensure proper mixing for special parameters as well)
Mixing is one of the many elements that affect how accurate the results are. As different types of chemistry demand different types of mixing; there CXL Pro Plus stands alone with a triple-speed mixing mechanism. Triple speed mixing mechanism ensures the perfect mixture for a special parameter like CRP, ASO, RF, etc. This mechanism has High, Medium & Low speeds of mixing for different types of parameters.
Real-time water quality checks
Another big factor for the precise result is water quality. CXL Pro Plus indicates the water quality before measuring the samples. By performing this function, CXL Pro Plus ensures tests with optimum conditions and saves reagents. No result variation or recalibration is required due to water contamination.
Special degasser device to eliminate the air & bubble
CXL Pro Plus has vertical collision protection, which prevents probe breakage. The degassing device eliminates the bubble formation. The digital liquid level detection keeps track of reagent consumption. The flat paddle stirrer mixing ensures proper homogeneous mixing. Mispa CXL Pro plus offers the best sample/ reagent pipetting mechanism.
On-board laundry system with special reverse function
CXL Pro Plus offers 7 stops and 11 steps of an efficient laundry system. Quality of cleaning is checked after every step during washing, ensuring zero carryover. The reverse washing function of the probe makes the system unique and ensures the best onboard laundry system which results in zero carryover.
HbA1C with whole blood
HbA1c is a test by which we can detect the perfect sugar level. To perform HbA1c in a conventional fully automated biochemistry analyser, the operator needs to lyse the whole blood manually where there are chances of several manual errors. CXL Pro Plus offers automated-on board lysing, giving you the feel of ultimate automation.
Minimal water & detergent consumption
Mispa CXL Pro plus has low operational cost in terms of water consumption and detergent consumption. Only 4 Ltr/hr on performing 240 tests made the instrument the ultimate solution.
In terms of cuvette washing, 40 microliters concentrated wash is the only need to wash each cuvette. It ensures the longer and optimum usage of the wash solution. The single wash solution is the modern-day requirement that has been taken care of in Mispa CXL Pro Plus.
Mispa CXL Pro Plus does the best possible clinical chemistry analysis and has the best features-
-
Mispa CXL Pro Plus has a multi-functional reagent and sample carousel.
-
Mispa CXL Pro Plus has up to 80 positions of reagents/samples.
-
Mispa CXL Pro plus has a pre and post-dilution facility of samples which gives better flexibility on the usage of a single sample for multiple uses.
-
Mispa CXL Pro Plus has a reagent volume of 10~300 µl and a sample volume of 2~35 µl. The minimum amount ensures the minimal Cost per test.
-
Mispa CXL pro plus have a semi-permanent reaction cuvette, which ensures the best uses & best result accuracy for routine chemistry and special parameters. The 56 positions make it more compact.
-
Mispa CXL Pro plus have a minimum reaction Volume of 100 µl.
-
Mispa CXL Pro Plus maintains the best Reaction disk constant temperature with a thermostat air bath. It saves the extra water consumption as well.
-
Mispa CXL Pro Plus has a calibration method of Linear (One-point, two-point and multipoint), Logit-Log 4P, Logit-Log 5P, Spline, Exponential, and Polynomial.
-
Mispa CXL Pro Plus Quality control method of Real-time, daily, monthly & QC chart | Visual QC assessment | QC history checking QC error analysis.
Physical Characteristics of Mispa CXL Pro Plus
|
Characteristic |
Details |
|
Dimensions |
744 x 703 x 530 (L x W x H) |
|
Weight |
100 Kg (approx..) |
|
PC Operating System |
Windows 7 or Windows 10 |
|
PC Configuration |
CPU > 2.9 GHz (dual-core processor); RAM > 4 GB; Harddisk ≥ 160 GB |
|
Power Supply |
100- 240 VAC, 50/60 Hz |
|
Power Consumption |
600 V |
|
PC interface |
RS 232 Port/ TCP/IP network connection, standard RJ-45 |
Technical Specification of Mispa CXL Pro Plus
|
Characteristic |
Details |
|
Photometric System |
HCFG rear Spectrophotometry |
|
Measuring Principle |
Spectrophotometry |
|
Wavelengths |
Total 12 Wavelengths (340nm, 380nm, 405nm, 450nm, 480nm, 505nm, 546nm, 570nm, 600nm, 660nm, 700nm, 750 or 800nm) |
|
Methodology |
Endpoint, Fixed time, Kinetic, Single & Dual reagent chemistries Mono & Bi chromatic, Linear & non-linear multipoint calibration |
|
Photometric Throughput |
Up to 240 tests/hour |
|
Light Source |
Halogen - Tungsten lamp (12V/20W) |
|
Reagent/Sample position |
Upto 80 Positions |
|
Sample/Reagent Probe |
Digital liquid level detection and vertical collision protection. |
|
Reagent Volume |
10~300 µl |
|
Sample Volume |
2~35 µl |
|
Reagent bottle volume |
20 & 70 ml |
|
Sample Dilution |
Pre & Post dilution facility |
|
Reaction Cuvette |
56 positions optical plastic cup |
|
Reaction Volume |
100~360 µl |
|
Reaction Temperature |
37±0.1°C |
|
Mixing System |
Teflon coated stirrer with triple speed mixing mechanism |
|
Reaction disk constant temperature |
Thermostat air bath |
|
Laundry system |
Efficient system adopting 7 stops, 11 steps |
Safety and Maintenance Protocol of Mispa CXL Pro Plus
Periodic maintenance of the unit is to be performed by the service engineer of Agappe.
-
Mispa CXL Pro Plus is for indoor use only.
-
The Installed site should be free of dust and away from direct sun.
-
The site should be free of corrosive and inflammable gas.
-
The bearing or the ground should be free from vibration
-
Cover the instrument with the dust cover when not in use.